B&K Precision 1856D - Manual User Manual

Page 7

Advertising
background image

7

(Table 1)

 Max. Input Voltage Level

FIG. 1 MAX. Input Level. (Input A)

 PERIOD RANGE:

10 nS to 10 S

DISPLAY: n/u/m (nano/micro/milli) Sec. with dec point

Accuracy: + Time base stability + 0.3% + 1 count

 TOTAL RANGE:

10 Hz to 30 MHz

CAPACITY: 0 to 999 999 999,

OVER FLOW : “OF”

 RPM RANGE:

600 to 600 x 10

6

RPM, OVER FLOW : “OF”

 INPUT C. CHARACTERISTICS

 FREQUENCY RANGE: 80 MHz to 3.5 GHz
 SENSITIVITY:

15mV from 80 MHz to 2000 MHz

20mV from 2000 MHz to 3.0 GHz

30mV from 3.0GHz to 3.2GHz

50mV from 3.2GHz to 3.5GHz

 COUPLING:

AC only

 IMPEDANCE:

50

±5%

 MAX. INPUT LEVEL: 3 V rms sine wave
 Accuracy:

+ Time base stability + resolution (table 2) + 1 count

300

250

150

200

100

25

50

MA

X

INP

UT

V

OLT

A

GE

(AC

V &

D

C

V)

5

0

100

500

FREQUENCY (Hz)

1K

1M

5M

10M

50M

100M

Advertising