12 IOM-DPCDPR-A: FEB 2014

5.2. DPC...- Measurement ranges and metrological parameters.

5.2.1. DPC..., Measurement ranges

N

Nominal measuring range

(FSO)

Minimum set range

Rangeability

Overpressure limit

(without hysteresis)

1.

0...1000bar

(0...100MPa)

10bar

(1MPa)

100:1

1200bar (120MPa)

2.

0...300bar

(0...30MPa)

3bar

(300kPa)

100:1

450bar

(45MPa)

3.

0...160bar

(0...16MPa)

1,6bar

(160kPa)

100:1

450bar

(45MPa)

4.

0...70bar

(0...7MPa)

0,7bar

(70kPa)

100:1

140bar

(14MPa)

5.

0...25bar

(0...2,5MPa)

0,25bar

(25kPa)

100:1

50bar

(5MPa)

6.

0...7bar

(0...0,7MPa)

0,07bar

(7kPa)

100:1

14bar

(1,4MPa)

7.

-1...6bar

(-100...600kPa)

300mbar

(30kPa)

23:1

14bar

(1,4MPa)

8.

0...2bar

(0...200kPa)

100mbar

(10kPa)

20:1

4bar

(400kPa)

9.

0...1bar

(0...100kPa)

50mbar

(5kPa)

20:1

2bar

(200kPa)

10.

-0,5...0,5bar (-50...50kPa)

50mbar

(5kPa)

20:1

2bar

(200kPa)

11.

0...0,25bar

(0...25kPa)

25mbar

(2,5kPa)

10:1

1bar

(100kPa)

12.

-100...100mbar (-10...10kPa)

20mbar

(2kPa)

10:1

1bar

(100kPa)

13.

-15...70mbar* (-15...7kPa)

5mbar

(0,5kPa)

17:1

0,5bar

(50kPa)

14.

-7...7 mbar* (-700...700Pa)

1mbar

(0,1kPa)

14:1

0,5bar

(50kPa)

15.

0...1,3bar abs (0...110kPa abs)

50mbar abs (5kPa abs)

26:1

2bar

(200kPa)

16.

0...7bar abs

(0...7MPa abs)

0,07bar abs (7kPa abs)

100:1

14bar

(1,4MPa)

17.

0...25bar abs (0...2,5MPa abs)

0,25bar abs (25kPa abs)

100:1

50bar

(5MPa)

18.

0...70bar abs (0...7MPa abs)

0,7bar abs (70kPa abs)

100:1

140bar

(14MPa)

* - only for transmitters without diaphragm seal,

5.2.2. DPC..., Metrological

parameters

Accuracy

max ± 0,075% for the basic range (0,16% for range 14)

Long term stability

accuracy for 3 years (for the nominal measuring range)

Error due to supply voltage changes

max ± 0,002%(FSO)/1V

Thermal error

max ± 0,08%(FSO)/10ºC

max ± 0,1% FSO/10

C for n°12, 13, 14 ranges.

Thermal error for the whole thermal

max ± 0,25%(FSO)

compensation range

(max ± 0,4% FSO/10

C for n°12, 13, 14 ranges.

5.2.3. DPC..., Pressure Connectors

M-type connector with M20x1.5 thread

– see figure 5a,

P-type connector with M20x1.5 thread

– see figure 6a,

CM30x2-type connector with flush diaphragm

– see figure 7a,

G1/2 -

type connector with G1/2” thread – see figure 8a,

GP -

type connector with G1/2” thread,

CG1-

type connector with G1” thread and flush diaphragm – see figure 8e,

RM-type connector with M20x1.5 thread and radiator
RP-type connector with M20x1.5 thread and radiator