Troubleshooting guide – Dynasonics UFX ULTRASONIC FLOW METER User Manual

Page 12

Advertising
background image

12

TROUBLESHOOTING GUIDE

Unit does not turn “ON”
when ON/OFF key is
pressed

•

Verify that batteries are installed and
contain a charge.

“Err1” is indicated

•

The batteries must be replaced.

No display readings are
obtained and no “OK”
icon is observed

•

Poor acoustic coupling to pipe. Apply
silicone grease to transducer.

•

Ensure pipe is full of a flowing liquid.

•

If the pipe has a plastic liner, move the
transducer to another location. The liner
may contain an air void.

•

Non-working transducer. Rub
transducer head with thumb for reading.

•

Liquid contains less than 3% total
suspended solids.

•

Move transducer closer to a source of
flow disturbance (i.e. an elbow, pump
outlet or control valve).

Dashes appear on display

•

Liquid velocity is greater than 30 FPS (9
MPS).

Readings are obtained,
but the “OK” icon does
not turn on.

•

Signal strength is low. Flow readings
may be erroneous. Move the transducer
closer to a source of hydraulic
disturbance.

Erroneous Readings

•

Transducer mounted incorrectly.

•

Another local ultrasonic instrument is
operating at approximately the same
frequency as the UFX.

•

Excessive pipe vibration.

•

Very viscous (thick) liquids will cause
the unit to read lower than actual flow.

•

The pipe is not completely full of liquid.