Quick-start operating instructions, Startup connections – Dynasonics 701_771C Series Insertion Ultrasonic Flow Meter User Manual

Page 3

Advertising
background image


Rev. 12/97 -2- D77X

Insertion Depth =

(pipe I.D. inches) X 0.125 + 0.67 inches

or

(pipe I.D. mm) X 0.125 + 17 mm

B. Secure the probe insertion distance using the upper

and lower brass jam nuts located on the threaded
rods.

C. Loosen the set screw located in the probe retaining

collar with the enclosed allen wrench.

D. Rotate the probe body till the arrowhead, located

on the top of the probe, is parallel with the pipe and
points down stream in the direction of flow.

E. Lock the probe rotation with the set screw in the

retaining collar.

F. Route the transducer cable back to the transmitter,

avoiding locations near high voltage supply wires.
Transducer connections are made through the left
conduit hole in the bottom of the transmitter.

3. TRANSDUCER CONNECTION

A. DO NOT attempt to connect additional cable to the

factory supplied transducers.

B. Connect the transducer spade terminals to the

XDCR terminal block located on the lower left
corner of the main PCB.

4. INITIAL SETTINGS AND POWER UP

A. Route a stable, grounded power supply to the

transmitter. Power wiring is made through the right
conduit hole.

B. Connect Power to the upper right terminal block as

appropriate. Apply power.

C. Press the RESET key on the keypad.

D. If the pipe is full of a flowing liquid, the default
readings will be FPS (Feet per Second).

Startup

Connections

QUICK-START OPERATING INSTRUCTIONS