Part 2 - transducer installation, Table 2.1, Transducer mounting locations – Dynasonics D300_301 Series Doppler Ultrasonic Flow Meter User Manual
Page 10
Advertising

Rev. 2/00 -2.2- 300/301
The following list outlines how to install the DT6
transducer for optimal performance, highest reliability
and greatest accuracy:
1. Select a transducer site at least 10 pipe diameters
downstream from bends, or fittings and 5 pipe
diameters upstream. A symmetrical flow pattern is
necessary for accuracy and repeatability over the
PART 2 - TRANSDUCER INSTALLATION
Table 2.1
1
1
The D300/301 system will provide repeatable measurements on piping systems that do not meet
these requirements, but the accuracy of these readings may be influenced to various degrees.
Transducer Mounting
Locations