OWNER’S MANUAL

Diagrammatic Abator Instructions:

Installing Optional New Abator Cartridge, Resetting Counter

When the Abator cartridge has been used completely
(250 sterilization cycles), the screen will display
ABATOR CARTRIDGE EXPIRED and the alarm will
sound.

Additional cycles may be completed by bypassing the
screen. This is not recommended. The cartridge
should be changed prior to starting a new cycle.

ABATOR CARTRIDGE

EXPIRED

EXIT BYPASS

Once a new cartridge is installed, the
counter must be reset.

To reset the counter:
1. Hold the left button continuously for
more than 3 seconds










2. Press ABATOR button





3. Answer YES to the question:
INSTALL NEW ABATOR
CARTRIDGE

When

left button

pressed and held

for more than 3

seconds.

AN74 I

ANPROLENE STERILIZER

PARAMETERS START

INSTALL NEW

ABATOR CARTRIDGE

NO YES

PRESS ABATOR TO

INSTALL CARTRIDGE,

OTHERWISE PRESS DIAG

ABATOR DIAG

CARTRIDGE COUNTER

RESET TO 250

DIAGNOSTICS

ABATOR ATTACHED

© Andersen Products Ltd