Mounting with a Stand‐Pipe:  
A stand‐pipe may  be used to  dampen turbulence, separate  surface foam from the point of 
measurement  or  increase  performance  in  heavy  vapor.    When  mounting  the  sensor  in  a 
stand‐pipe, the minimum diameter of the pipe is 2”.  Larger diameter pipes can be used.  The 
pipe  should  be  attached  with  a  coupling  and  reducer  bushing.    The  pipe  length  should  run 
the measurement span and the bottom of the pipe should remain submerged at all times to 
prevent  foam  from  entering  the  pipe.    Cut  the  bottom  end  of  the  pipe  at  45°  and  drill  a 
1/4”pressure  equalization  hole  high  in  the  dead  band.    The  pumps  should  not  drive  liquid 
past the open end of the standpipe which causes the liquid in the pipe to oscillate.  

Stand‐Pipe Mounting

IMPORTANT MOUNTING GUIDELINES: 

1) Never mount the sensor at an angle. 

2) Liquid should never enter the dead band. 

3) Mount at least 3” from the side wall. 

4) Never mount in a vacuum. 

5) Do not obstruct the sensor’s beam width. 

 

 