General information, Safety information – Hach-Lange POLYMETRON 9240 Basic User Manual User Manual

Page 4

Advertising
background image

Specification

Details

Measurement category

Cat II, Class 1 (overvoltage < 1500V)

Maximum panel dimension (H x L x
D)

850 x 450 x 252.5mm [33.46 x 17.71 x 9.94in]

Inlet

Simple fittings for 6 mm O.D. tubing or ¼" O.D. in PE-low density. ¼" OD
in PHED-PTFE-SS as option

Outlet

Barbed stem for 12 mm (½" I.D.) hose

Protection rate

Transmitter: IP65 (NEMA 4)
Panel: IP50 (dust protection)
Optional Enclosure: IP54 (splash water proof)
Instrument is designed to avoid DIPA vapor inside the enclosure. All DIPA
vapor is collected and sent to the instrument drain

Cell

PMMA - compact (minimum tubing)

Flame rate

Conform UL

Maximum weight

15 - 30 Kg

Mains power supply

100 - 240 VAC, 50-60 Hz, ± 10%, automatic switching

Max. consumption

80 VA

Fuse

5 x 20 cartridge T2AL-250V following CEI127

Display

Curve trend, Last Cal Date, Historical, Concentration, Temperature,
Potential

Analog outputs

Number: 6
4-20 or 0-20 mA (650 ohms)
Linear / Dual / Logarithm
Event indication

Relays

4 x Relay (conc)
1 x Warning
1 x System

Logic input

Active / Inactive channels
Remote AutoCal

European standards

EN 61326 Class A for EMC; EN 601010-1 for low voltage safety

International standards

cETLus

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.

Safety information

N O T I C E

The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.

4 English

Advertising