LDO™ Sensor Replacement for LDO101 Probes

Each LDO sensor cap replacement kit comes with a sensor-specific
calibration update iButton

®

. When the iButton is installed in the pressure-

sensor module (located on the LDO101 probe cable), the probe calibration is
updated for the new sensor cap.

The kit includes:

Sensor cap with O-ring and seal

Calibration update iButton

Replacing the Sensor Cap and iButton

1. Disconnect the LDO101 probe from the meter. If using a rugged LDO101

probe, remove the protective shroud from the end of the probe.

2. Remove the used sensor cap from the probe and discard. Do not touch or

soil the exposed sensor lens (

Figure 1

).

3. Carefully remove the used O-ring from the sensor tip. Remove the used

cap seal by rolling it off with your fingers. Do not use sharp metal tools.