Quick Start Guide

For si792 D series 2-wire transmitters

DOC016.53.00797

Additional Configuration

Note: Additional configuration is for 4-20mA /Hart version only. For Foundation Fieldbus or Profibus PA,

refer to the user manual for configuration instructions.

Refer to

Table 4

for default passcode settings.

1. To enter the configuration mode, press

CONF

, configuration code:

1200

,

ENTER

. There are 5

configuration menus:

• current output "out.1MNU"
• temperature compensation "tc MNU"
• corrections configuration "Corr MNU"
• Calibration setup "Cal MNU"
• alarm settings "ALrt MNU"

2. In configuration mode, press the

RIGHT ARROW KEY

to navigate to a specific configuration

menu. Press

ENTER

to access the menu options.

3. Use the

ARROW KEYS

to edit values. Press

ENTER

to save the settings. If a value is outside of

the acceptable range, “Err” will be displayed and the value will not be accepted.
4. To exit the configuration mode and return to measurement mode, press

CONF

,

ENTER

at

any time.

Refer to user manual for detailed configuration instructions and a complete list of menu options.

Table 4 Default passcode settings

Display

Function

Default setting

Cal INFO

Calibration information

0000

CAL 0-CAL

Calibration zero

1001

CAL CAL

Calibration Saturation/Concentration

1100

CAL PROD

Product calibration

1105

CAL RTD

Temperature sensor adjustment

1015

CFG ERR

Error information

0000

CFG CONF

Configuration

1200

CFG SNSR MO Probe monitor

2222

CFG OUT

Current source

5555

CFG SPCL ST

Administrator passcode. Use

ARROW

keys to

select NO or YES

1989

NO SPCL ST

Do not change
administrator passcode

Press

ENTER

to return to the default

passcode
Press

CONF

to exit the editor

YES SPCL ST

Change administrator
passcode

Press

ENTER

to accept the new

passcode
Press

CONF

to exit the editor