Do not use variable volume, Flow based collection, Use counts for the flow source – Hach-Lange SD900 Basic User Manual User Manual

Page 20: Use 4‍–‍20 ma input for the flow source, Restore the default settings, Flow based—refer to, If no, refer to

Advertising
background image

Option

Description

Collection period

Specify the total collection period in hours and minutes (0:01–999:00).

Minimum sample volume If the flow volume is not sufficient to collect this minimum volume, the sample is

skipped (10–10,000 ml).

Do not use variable volume

1. Select or enter the values for the option.

Option

Description

Take first sample Select whether the program starts immediately or after the first time interval has passed.

Flow based collection

Specify the flow source for flow based sampling:

Counts—Refer to

Use counts for the flow source

on page 20.

• 4–20 mA input—Refer to

Use 4–20 mA input for the flow source

on page 20.

Use counts for the flow source

1. Select or enter the values for each option.

Option

Description

Take sample every Collect samples at regular flow intervals, in counts (1–9999 counts).

Overrride time

Collect a sample if the flow volume is very low.

Time

The maximum time between samples (0:01–999:00). The timer starts again after each
sample is collected.

Take first sample

The program starts immediately or after the first time interval.

Use 4–20 mA input for the flow source

1. Select or enter the values for each option.

Option

Description

Select unit

Selects the flow units to use.

Map 4 mA input

Enter the flow rate that would agree with the 4 mA input.

Map 20 mA input

Enter the flow rate that would agree with the 20 mA input.

Take sample every Enter the total flow after which the sample has to be drawn.

Override time

Select enable to push a sample to be collected if the flow volume is unusually low.

Time

The maximum time between samples (0:01–999:00). The timer starts again after each
sample is collected.

Sample volume

Enter the volume to be collected per sample. If bottles/sample mode is selected, each
bottle receives a full sample volume (100–10,000 mL). This option is skipped if variable
volume is selected.

Restore the default settings

The restore option sets all of the program parameters to the default settings.

20 English