N O T I C E

Potential instrument damage. Delicate internal electronic components can be damaged by static electricity,
resulting in degraded performance or eventual failure.

To install the module and connect the sensor, refer to the illustrated steps that follow and the wiring
table

Table 1

.

Be sure to connect all sensor ground/shield wires to the controller enclosure grounding screws.

Note: If the sensor cable is not long enough to reach the controller, an interconnect cable and junction box are
required to extend the distance.

Table 1 Dissolved oxygen and oxygen scavenger sensor wiring

Terminal

Description

Dissolved oxygen

Oxygen scavenger

1

Temp +

Black

Black

2

Temp -

Blue

Blue

3

4

5

Ground

Green

Green

6

Earth

Yellow

Yellow

7

8

9

Working electrode

White

White

10

Counter electrode

Red

Gray

11

Reference electrode

Transparent

12

In Step 3 of the following sequence, set the switch as follows:

Switch Setting

Description

Setting 1

Dissolved oxygen

Setting 2

Oxygen scavenger