3 measurement board, 8 measurement alarm relays, Installation 3.7.3 measurement board – Hach-Lange ORBISPHERE K1100_M1100 Operator Manual User Manual
Page 36
Advertising

34
Installation
3.7.3 Measurement board
3.8 Measurement alarm relays
Figure 17 Measurement board
Figure 18 Connector J7
Connector J7 (inputs & outputs)
Measurement alarms relays:
1.
Common
2.
Output relay 1
3.
Output relay 2
4.
Output relay 3
Analog current outputs:
5.
GND
6.
Output 1
7.
Output 2
8.
Output 3
Digital inputs:
9. to 16.
Not used
Figure 19 Output relays
The three output relays are located on the measurement
board.
They can be individually configured to Normally Open (NO) or
to Normally Closed (NC) by physically moving the jumper on
each relay. On the picture example on the left:
•
Upper relay is set to NC
•
Middle relay is set to NO
•
Lower relay is shown with no jumper
Note:
J14 (upper relay) is relay 1
J18 (middle relay) is relay 2
J19 (lower relay) is relay 3