Approx. constant

10 cm

-1

Measuring range

100 µS/cm to 500 mS/cm

Operating temp.

0 to 80 °C

Max. pressure

2 bar (25 ºC)

Temperature sensor

Pt 1000

Body material

glass

Electrode material

platinum

Thread

PG 13,5

Connector

MP-5

Min. immers. depth

50 mm

5

GB

Storage
Store the cell dry.

Life of the cell
The cell lifespan is unlimited if suitable
maintenance is applied and the cell does
not break.

Troubleshooting
... Reading 0 independently of the meas-
ured sample.

Broken cable.
Replace the cell.

... Unstable reading.
... Impossible calibration.

Dirty measuring electrodes or air
bubbles on their surface.
Clean or shake slightly the cell.

... The obtained readings differ a lot from

the real values.

Verify that the conductivity is covered
by the measuring interval of the cell.
Re-calibrate using standards in
good conditions.

.... slow cell response

Regenerate the cell.

Specifications

Applications
Samples with high conductivities, sea wa-
ters, industrial process waters, bleaches
and concentrated acids, etc.

QUALITY CERTIFICATE

The attached cell 53 99 has passed the
quality test and fulfils the following speci-
fications:
• EC error 5 %
between 100 µS/cm and 500 mS/cm.
• Temperature error 0.5 ºC.

53 99