Hach-Lange LZU5396 User Manual

Page 5

Advertising
background image

Approx. constant

1 cm

-1

Measuring range

1 µS/cm to 200 mS/cm

Operating temp.

0 to 80 °C

Max. pressure

2 bar (25 ºC)

Body material

glass

Electrode material

platinum

Thread

PG 13,5

Connector

S8

Min. immers. depth

15 mm

5

GB

Storage
Store the cell dry.

Life of the cell
The cell lifespan is unlimited if suitable
maintenance is applied and the cell does
not break.

Troubleshooting
... Reading 0 independently of the meas-
ured sample.

Broken cable.
Replace the cell.

... Unstable reading.
... Impossible calibration.

Dirty measuring electrodes or air
bubbles on their surface.
Clean or shake slightly the cell.

... The obtained readings differ a lot from
the real values.

Verify that the conductivity is covered
by the measuring interval of the cell.
Re-calibrate using standards in
good conditions.

.... slow cell response

Regenerate the cell.

Specifications

QUALITY CERTIFICATE

The attached cell 53 96 has passed the
quality test and fulfils the following speci-
fications:
• EC error 5 %
between 1 µS/cm and 200 mS/cm.
• Temperature error 0.5 ºC.

53 96