Hach-Lange LZU5395 User Manual
Page 5
Approx. constant
0.1 cm
-1
Measuring range
0.1 µS/cm to 50 mS/cm
Operating temp.
0 to 80 °C
Max. pressure
2 bar (25 ºC)
Body material
glass
Electrode material
platinum
Thread
PG 13,5
Connector
S8
Min. immers. depth
45 mm
5
GB
Storage
Store the cell dry.
Life of the cell
The cell lifespan is unlimited if suitable
maintenance is applied and the cell does
not break.
Troubleshooting
... Reading 0 independently of the meas-
ured sample.
Broken cable.
Replace the cell.
... Unstable reading.
... Impossible calibration.
Dirty measuring electrodes or air
bubbles on their surface.
Clean or shake slightly the cell.
... The obtained readings differ a lot from
the real values.
Verify that the conductivity is covered
by the measuring interval of the cell.
Re-calibrate using standards in
good conditions.
.... slow cell response
Regenerate the cell.
Specifications
QUALITY CERTIFICATE
The attached cell 53 95 has passed the
quality test and fulfils the following speci-
fications:
• EC error 5 %
between 0.1 µS/cm and 50 mS/cm.
• Temperature error 0.5 ºC.
53 95
53 95_DOC013.97.90292_Ed 0510.indd 5
3 95 DOC013 97 90292 Ed 0510 i dd 5
20/9/10 16:17:29
20/9/10 16 17 29