Hach-Lange LZU5390 User Manual

Page 5

Advertising
background image

Approx. constant

0.15 cm

-1

Measuring range

0.1 µS/cm to 10 mS/cm

Operating temp.

0 to 100 °C

Max. pressure

6 bar (25 °C)

Temp. sensor

Pt 1000

Body material

titanium, PVDF, s.steel

Electrode material

titanium

Thread

1/2” NPT

Connector

MP-5

Min. immers.
depth

10 mm

5

GB

Life of the cell
The cell lifespan is unlimited if suitable
maintenance is applied and the cell does
not break.

Troubleshooting
... Reading 0 independently of the mea-

sured sample.

Broken

cable.

Replace the cell.
... Unstable reading.
... Impossible calibration.
Dirty measuring electrodes or air

bubbles on their surface.

Clean or shake slightly the cell.
... The obtained readings differ a lot from

the real values.

Verify that the conductivity is covered

by the measuring interval of the cell.

Re-calibrate using standards in good

conditions.

... slow cell response
Regenerate the cell.

Specifications

Applications
Clean waters, low conductivity waters,
etc.

CERTIFICATE OF QUALITY

The attached cell 53 90 has passed the
quality test and fulfils the following speci-
fications:
• EC error 5 %.
between 0,1 µS/cm and 10 mS/cm.
• Temperature error 0.5 ºC.

53 90

53 90_DOC013.97.90290_Ed 0510.indd 5

3 90 DOC013 97 90290 Ed 0510 i dd 5

20/9/10 16:16:15

20/9/10 16 16 15