If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.

7.

Push

EXIT

until the meter returns to the measurement mode.

Calibration reminder

Reminder repeat: Off, 2 h, 4 h, 8 h, 2 d, 5 d, 7 d
Expires: Immediately, Reminder + 30 min,
Reminder + 1 h, Reminder + 2 h, Continue Reading

Options for calibration reminders

Standard units

(For custom calibration standard only)

Set units for custom calibration standard

Standard value

(For custom calibration standard only)

Set values for custom calibration
standard

Reference temperature

(For custom calibration standard only)

Set reference temperature for custom
calibration standard

Temperature correction

(For custom calibration standard only)

Set temperature correction for custom
calibration standard

Table 4 Calibration option settings

Settings

Options

Description

Troubleshooting

Message or symptom

Possible cause

Action

Probe not supported

Software not updated

Update the HQd software to the newest revision at

www.hach.com/SoftwareDownloads

.

HQd meter does not support IntelliCAL
probe.

Contact a Hach Technical Support Representative.

Connect a probe or
Probe requires service

Conductivity cell not connected properly

Disconnect, then connect the probe. Tighten the locking
nut.

Software not updated

Update the HQd software to the newest revision at

www.hach.com/SoftwareDownloads

. Refer to the HQd

Series meter manual.

Large number of methods stored on
probe.

Continue to let probe connect. Do not disconnect probe.

Damaged probe

Verify connectivity with another probe or meter to confirm
isolated issue with probe. Contact a Hach Technical
Support Representative.

Slow response time

Mineral or sample buildup on electrodes

Clean the probe as described.

Out of range

Temperature and/or pressure sensor
error

Verify that both temperature and pressure sensors are
reading accurately

Damaged probe

Replace the conductivity probe or contact a Hach
Technical Support Representative.

Sample outside of specifications

Verify that the sample concentration and temperature are
within the range of the CDC401.

Drifting/Inaccurate
readings

Incorrect settings

Measurement options: verify that Temp Correction
(Correction Factor if not set as NaCl) and reference
temperature are correct.

Calibration options: verify that Standard Value, Reference
Temperature and Temp Correction are correct.

Mineral or sample buildup on cell

Clean the probe as described on page

5

.

CO

2

absorption in LIS/high purity

samples

Isolate LIS/high purity samples to prevent sample
contamination.

Calibration failed - Out
of limits/Out of range

Incorrect settings

Measurement options: verify that Temp Correction
(Correction Factor if not set as NaCl) and reference
temperature are correct.

Calibration options: verify that Standard Value, Reference
Temperature and Temp Correction are correct.