Message or symptom

Possible cause

Action

Drifting/Inaccurate readings

Contaminated glass bulb

Clean the probe. Refer to

Clean the probe

on page 10.

CO

2

absorption (for low ionic

strength or high purity sample)

Use the LIS chamber for LIS/high purity samples to
prevent sample contamination.

Clogged reference

Make sure that the filling hole is not covered. Rinse the
reference junction holes with deionized water. Carefully
tap the probe with hand or shake the probe to remove air
bubbles. If there are salt crystals around the junction,
soak the electrode tip in a 60 ºC (140 ºF) water bath for
10 minutes.

Improper storage conditions/
Dehydrated glass bulb

Clean or condition the probe and calibrate again. The
probe can not operate correctly if the probe was left dry
for a long time. Rehydrate the probe. Refer to

Storage

on page 11.

Colloidal and/or particulates in
the filling solution

Replace the filling solution, calibrate and test again.

Air bubbles around inner
reference electrode

Carefully tap the probe with hand or shake the probe to
remove air bubbles.

Not sufficient electrolyte flow

1

Open the sleeve junction.

Not sufficient stirring

Stir the standards and samples at a slow and steady rate
to prevent the formation of a vortex.

Out of range

Measurement value is out of
range

Make sure that the sample is in the range of the probe.

Out of limits

Check Standard value is out of
limits in the current method

Make sure that the standard is in the limits of the current
method.

Do another method that expands the acceptable limits.

Measurement value is out of
measurement limits in the current
method

Make sure that the sample is in the limits of the current
method.

Make a new method with an expanded range.

Temperature out of range

Temperature value is out of
range

Make sure that the sample temperature is in the range of
the probe.

Make sure that the temperature sensor works correctly.

Measured temperature is out the
range of the probe

Make sure that the standard temperature is in the range
of the probe.

Make sure that the temperature sensor works correctly.

Check standard temperature
value is out of range

Make sure that the check standard temperature is in the
range of the probe.

1

Only applicable for probes with sleeve junction.

Consumables and accessories

Description

Quantity

Item Number

Electrode filling solution, saturated KCl

59 mL

25118059

Electrode filling solution, saturated KCl

500 mL

25118049

KCl crystals (Radiometer Analytical)

15 g

S21M001

Hach Electrode Cleaning Solution

500 mL

2965249

EDTA standard solution, 0.20N

500 mL

102149

Soaker bottle

1

5192900

14