Message or symptom

Possible cause

Action

Temperature out of range

Calibration temperature value is
outside of range

Make sure that the sample temperature is within the
range of the probe.

Make sure that the temperature sensor is working
correctly.

Measured temperature is outside
the range of the probe.

Make sure that the standard temperature is within
the range of the probe.

Make sure that the temperature sensor is working
correctly.

Check standard temperature value
is outside of range

Make sure that the check standard temperature is
within the range of the probe.

Below detection limit

Measurement is not quantifiable
with current saved calibration
(based on IUPAC-defined practical
detection limit).

Perform a new calibration. Check that sample
concentration is bracketed between two standard
solution values (if within linear range).

Re-run calibration and measurement, optimizing
meter settings for slope acceptance and stabilization
criteria for expected sample concentration.

Re-run calibration and sample measurement with the
tips for low-level measurement.

Measurement value is outside of
range.

Make sure sample is within the range of the probe.

Check probe response
To make sure there is a probe response, measure the probe potential (in mV) of two
Ammonia Standard Solutions one decade apart that are above and below the expected
sample concentration. For example, use 10 and 100 mg/L Ammonia Standard Solutions.
The two solutions should have potentials (difference in mV readings) that are 57 mV apart
at 25 ºC (within the slope limits of the method is acceptable). Both solutions should be
above 0.5 mg/L Ammonia.

Check accuracy of sample reading
To make sure the sample measurement is accurate, add a spike of Ammonia Standard
Solution with the volumetric pipet. Refer to

Table 3

and formulas to calculate the percent

of recovery.
Typically a percent of recovery of 100% ±5% is a good indication that the instrument,
technique and the sample do not contribute to measurement errors.

Table 3 Spike reference

Measured sample concentration

Volume of standard at add

Concentration of standard

1 to 2 mg/L

0.5 mL

100 mg/L

3 to 6 mg/L

1.0 mL

100 mg/L

7 to 15 mg/L

0.3 mL

1000 mg/L

15 to 30 mg/L

0.5 mL

1000 mg/L

30 to 60 mg/L

1.0 mL

1000 mg/L

Percent recovery
Use the following formula to calculate the percent recovery when the sample volume is
25 mL:
E = (C x V

1

) / V

2

R = (A / (E + S)) x 100

S = mg/L of Ammonia in sample (before spike)

C = concentration of standard used for spiking (mg/L)