Measuring range
± 2000 mV
Indicator metal
platinum
Operating temp.
0 to 80 °C
Max. pressure
6 bar
Reference element
encapsulated
Ag/AgCl crystals
Diaphragm
3 ceramics
Electrolyte
gel
Body material
glass
Thread
PG 13,5
Connector
S8
Min. immersion
depth
17 mm
5
2
GB
Specifications
Applications
Swimming pools, drinking water, gas
scrubbing towers, boiler waters, detoxifi -
cation, outlet waters from a waste water
treatment plant (WWTP), etc.
Limitations
Low-conductivity solutions.
Viscous or dirty solutions able to clog quickly
the diaphragm.
CERTIFICATE OF QUALITY
The attached electrode 53 62 has passed
the quality test and follows the next speci-
fications:
• Deviation < ± 10 mV in redox standard.
• Response time in redox standard < 20 s.