Hach-Lange 5362 User Manual

Page 5

Advertising
background image

Measuring range

± 2000 mV

Indicator metal

platinum

Operating temp.

0 to 80 °C

Max. pressure

6 bar

Reference element

encapsulated

Ag/AgCl crystals

Diaphragm

3 ceramics

Electrolyte

gel

Body material

glass

Thread

PG 13,5

Connector

S8

Min. immersion
depth

17 mm

5

2

GB

Specifications

Applications
Swimming pools, drinking water, gas
scrubbing towers, boiler waters, detoxifi -
cation, outlet waters from a waste water
treatment plant (WWTP), etc.

Limitations
Low-conductivity solutions.
Viscous or dirty solutions able to clog quickly
the diaphragm.

CERTIFICATE OF QUALITY

The attached electrode 53 62 has passed
the quality test and follows the next speci-
fications:
• Deviation < ± 10 mV in redox standard.
• Response time in redox standard < 20 s.

53 62_DOC013.97.90287_Ed 0510.indd 5

3 62 DOC013 97 90287 Ed 0510 i dd 5

23/9/10 10:21:56

23/9/10 10 21 56