Hach-Lange 5355 User Manual
Page 5

Measuring range
± 2000 mV
Indicator metal
platinum
Operating temp.
0 to 80 °C
Max. pressure
2 bar
Reference
element
encapsulated
Ag/AgCl crystals
Diaphragm
open circular
Electrolyte
solid
Body material
glass
Thread
PG 13,5
Connector
S8
Min. immersion
depth
12 mm
5
5
GB
Specifications
Applications
Dirty water, with mud, wastewater, etc.
Limitations
Rapid alteration of the polymer by: solutions
with pH < 2, very clean water, distilled and
de-mineralised water.
Temperature > 80ºC. Pressure > 2 bar.
This electrode is not recommended for instal-
lation on pipe.
CERTIFICATE OF QUALITY
The attached electrode 53 55 has passed
the quality test and follows the next speci-
fications:
• Deviation < ± 10 mV in redox standard.
• Response time in redox standard < 20 s.