Hach-Lange 5330 User Manual

Page 5

Advertising
background image

Measuring range

0 to 14 pH

Operating temp.

0 to 100 °C

Max. pressure

5 bar

Reference element

cartridge

Diaphragm

ring of porous PTFE

Electrolyte

gel

Body material

glass

Thread

PG 13,5

Connector

S8

Min. immersion
depth

12 mm

5

GB

... impossible to perform calibration.
... very slow response.

Dirt on the diaphragm or the mem-
brane.
De-hydrated membrane.
Electrode ageing or wear.

Clean, regenerate, replace the elec-
trode.

Storage
The connector should be protected from
humidity by the protective cap

.

Place the electrode in the storage protector

with KCl 3M.

Life of an electrode
The average life expectancy of a pH process
electrode oscillates between 6 or 12 months,
depending on the working conditions.
Electrodes exposed to high temperatures do
not last as long. Very alkaline mediums also
reduces the life span of electrodes.

If an electrode has too short life span,

less than three months, ask us about it.
Probably there is a more suitable elec-
trode for the application.

Specifications

Applications
Waste water, galvanising industry, osmosis
processes, demineralized water, dirty or
viscous samples , etc.

CERTIFICATE OF QUALITY

The attached electrode 53 30 has passed
the quality test and follows the next speci-
fications:
• Asymmetry potential < ± 15 mV.
• Sensitivity, pH 4 to 7 (at 25 ˚C) > 98%.
• Response time, pH 4 to 7 < 20 s.