Change calibration options – Hach-Lange ISECL18103 User Manual User Manual

Page 8

Advertising
background image

4. Select Measurement Options and update the settings:

Option

Description

Units

Sets the preferred unit for ISE measurements—mg/L (default), µg/L, g/L,
g/kg, mol/L, mmol/L, mol/kg, %, ppm or ppb.
Note: The mV units are shown when the detailed display is selected.

Significant Digits Sets the significant digits shown—2, 3 (default) or 4.

Auto
Stabilization

Sets auto stabilization—on (default) or off.
The default stability drift rate is 1.0 mV/min.

Stability Criteria

When Auto Stabilization is off, sets the stability criteria—0.1 to
9.9 mV/min.

Lower stability criteria will require longer stabilization times, but the
measurement will be more precise.

Higher stability criteria will require shorter stabilization times, but the
measurements may be less precise.

Measurement
Limits

Sets the measurement limits—Lower limit (default: 0.1 mg/L) or Upper
limit (default: 35,500 mg/L).
The measurement limits can be set to match the acceptable values for
the sample. When the measurement is above the upper limit setting or
below the lower limit setting, the meter shows an "Out of limits" message.
This message is an alert to a potential problem with the process
conditions.

5. If prompted, enter a name for the new method settings. Additional changes made to

the settings of an existing method are automatically saved with the same method
name.

6. Push EXIT until the meter returns to the measurement mode.

Change calibration options

1. Make sure a probe is connected to the meter.
2. Push and select ISECl181 Settings.
3. Select Modify Current Settings.
4. Select Calibration Options and update the settings:

Option

Description

Std Set

Sets the temperature compensated standard sets that are used for
calibration—

100, 1000 or 12500 mg/L

100 or 1000 mg/L

Standard set values are shown on the Calibration Options screen.
Custom standard sets are characterized at 25 °C (77 °F). Custom
standard values are not temperature compensated. Select the Custom
buffer to make a custom standard. Up to five standard values can be
made .
Note: Only the minimum calibration points must be measured for Done to

be shown on the calibration screen.

Chemical Form

Sets the chemical form.

Calibration
Units

Sets the preferred unit for ISE Calibration—mg/L (default), µg/L (available
only for custom calibration set), g/L, g/kg, mol/L, mmol/L, mol/kg, %, ppm
or ppb.

Std Set Values

When Std Set is set to Custom, sets the standard set values.
Up to five standard values can be made. Each standard value can include
a standard set value, Custom or No Standard.

8