Storage

The probe can be stored dry. To protect the sensing element, rinse with DI water and blot
dry with a lint-free cloth. Install the sensor protection cap.

Note: The probe must be conditioned after long-term storage. Refer to

Preparation for use

on page 2.

Troubleshooting

Message or symptom

Possible cause

Action

Probe not supported

Software not updated

To download the most current version of the
software, refer to the applicable product page on the
manufacturer's website.
Refer to the HQd Series meter manual for specific
instructions for the meter model.

HQd meter does not support
IntelliCAL

®

probe

Contact a Technical Support Representative.

Connect a probe or probe
requires service

Probe not connected properly

Disconnect, then connect the probe. Tighten the
locking nut.

Software not updated

To download the most current version of the
software, refer to the applicable product page on the
manufacturer's website.
Refer to the HQd Series meter manual.

Large number of methods stored
on probe.

Continue to let probe connect. Do not disconnect
probe.

Damaged probe

Make sure connectivity with another probe or meter
to confirm isolated issue with probe. Contact a
Technical Support Representative.

mV reading is the same for all
solutions

Electrical issue

Contact a Technical Support Representative.

Slow response time

Dirty sensing element

Clean the probe (refer to

Clean the probe

on page 10).

Low sample temperature or
temperature difference between
samples

Check the sample temperature. The lower the
temperature or the greater the difference of
temperatures between samples, traditionally the
longer the response time.

Bubbles trapped under sensor tip

Gently shake the probe until bubbles are removed
from under sensor tip.

Slope out of range (refer to

Check probe response

on page 13)

pH is incorrect

Make sure the pH is between 5.0 and 5.5 after each
ISA addition.

Ionic strength adjustor (ISA) not
used

Add ISA to each sample and standard (one powder
pillow per 25 mL of solution).

Insufficient conditioning

Condition for at least 30 minutes in a 0.1 mg/L
standard solution with ISA.

Damaged probe

Contact a Technical Support Representative.

Incorrect standards

Calibrate using freshly prepared standards.

Dirty sensing element

Clean the probe and recalibrate.

Bubbles trapped under sensor tip

Gently shake the probe until bubbles are removed
from under sensor tip.