Operation, Measurement, Changing temperature units – Hach-Lange SENSION+ MM110 User Manual

Page 13: Stability measuring, Continuous measuring, English, Once stability measuring has begun, press the, Key twice, Press the, Keys simultaneously

Advertising
background image

9

5. Operation

ENGLISH

Stable measurement

5.1. Measurement

Note: the instrument must be calibrated before reading
the first pH value.

1. Press

to turn on the instrument.

2. Remove the sensor protector.
3. Clean the sensor with deionized water.
4. Stir the sensor uniformly in the sample, holding it by

the top.

5.1.1. Stability measuring

Follow the diagram.
The reading is not stabilized on the screen until the
electrode signal varies less than 0.01 pH (0.5 mV) for
6 sec.
Once the reading has stabilized, the screen lights up
for 2 sec.

5.1.2. Continuous measuring

Once stability measuring has begun, press the

key

twice

.

The instrument displays the measured value directly on
screen at all times.
The continuous reading should always be started after
beginning a stability reading (see diagram).
To end a continuous reading, press

.

The display will light up for 2 sec.

Note: If the measuring instrument is not used for 5 minu-
tes, it shuts off automatically.

t

mV

t

mV

t

mV

t

mV

t

mV

t

mV

t

mV

t

mV

t

mV

2 s

+

Stability

measuring

Continuous

measuring

5.2. Changing temperature units

The instrument is shipped ready to measure temperature and
express the values in °C.
Follow the diagram to change the units to °F.

t

mV

t

mV

+

(1)

(1)

Press the

keys simultaneously.

sensION+ MM110 HACH_LANGE 0710.indd ANG:9

sensION+ MM110 HACH_LANGE 0710.indd ANG:9

7/9/10 15:32:45

7/9/10 15:32:45

Advertising