Assembly, Calibration, Maintenance – Hach-Lange CDC401-30 Instruction Sheet User Manual
Page 2

2
Assembly
Important Note: Do not remove the cable from the electrode. Doing so can damage the
probe.
Important Note: To ensure accurate temperature readings, do not cover the temperature
sensor with the shroud.
1. Attach a pair of color-code clips to the electrode and cable connector
(
Figure 1
, item
3
).
2. Rinse the probe with deionized water prior to use.
Calibration
For calibration steps, refer to the HQ Portable Meters User Manual. The IntelliCAL™
probes store the current calibration and calibration history in the probe memory. It is not
necessary to recalibrate when moving a calibrated electrode from one meter to another.
Maintenance
General Maintenance
•
Between samples, wipe the probe with a clean towel.
•
If the connection to the meter is poor, readings may be unstable. To ensure accurate
readings, check the connection to the meter and make sure the instrument is properly
grounded.
Storage
•
Between uses, store the probe dry in ambient conditions.