Assembly, Calibration, Maintenance – Hach-Lange PHC101-03 Instruction Sheet User Manual

Page 2

Advertising
background image

2

Assembly

1. Remove the electrode from the electrode storage cap (

Figure 1

, item

6

).

2. Remove the film that covers the two reference junctions (

Figure 1

, item

2

).

3. Rinse the electrode with deionized water and blot dry with a paper towel. Do not

scrub the bulb.

4. Attach a pair of color-code clips to the electrode and cable connector

(

Figure 1

, item

3

).

5. When ready to use, attach the electrode storage cap to the electrode storage cap lid

(

Figure 1

, item

6

) to prevent storage solution spillage and/or loss of the cap.

Calibration

For calibration steps, refer to the HQ Portable Meters User Manual. The IntelliCAL™
probes store the current calibration and calibration history in the probe memory. It is not
necessary to recalibrate when moving a calibrated electrode from one meter to another.

Maintenance

Important Note: Never use this electrode organic solvents; or if the pH to be measured is
lower than 2.

Important Note: Do not scrub the bulb. If the bulb becomes fouled, immerse and agitate
in a warm dilute mild detergent solution for a few minutes, rinse well with deionized water,
and gently blot or air dry prior to reuse.

General Maintenance

Before measuring a new sample, rinse the electrode with deionized water and blot dry
with a paper towel.

If calibration errors occur, recalibrate using fresh, high-quality buffers.

If the pH glass membrane is not properly hydrated, soak the electrode in pH 4 buffer
for up to two-hours.

Storage

Between uses, in intervals of up to two hours, store the electrode in the sample (if the
sample pH is not extreme), or in a neutral solution such as tap water.
For longer term storage, place the electrode in a 3M potassium chloride (KCl)
solution, either in the provided storage cap, or in a beaker.