Figure 2, Sensor installation in the flow-through unit – Hach-Lange NITRATAX sc Flow-through accessories User Manual

Page 3

Advertising
background image

3

Flow-through accessories installation for NITRATAX sc

Sensor installation in the flow-through unit

CAUTION
Pinch hazard. Only qualified personnel should conduct the tasks described in this
section of the manual.

Refer to

Figure 5

for the installation of the sensor in the flow-through unit.

1. Install the orientation screw to the sensor (

Figure 3

, item

3

or

Figure 4

, item

3

) to

make sure the measurement path in the measuring chamber is adjusted correctly.

Note: Replace the installed standard 50 mm sensor wiper with the supplied wiper with bore
wholes.

2. Insert the sensor (

Figure 5

,

Figure 6

) into the flow-through unit. Turn the sensor so

that the orientation screw slides into the cavity of the panel. Tighten the setscrew
(

Figure 3

, item

1

or

Figure 4

, item

1

) to the sensor.

3. Connect the inflow and the outflow tubes (

Figure 5

,

Figure 6

). Insert the inflow tube

approximately 50 mm into the screw connection of the 2 and 5 mm sensors.

4. Use the allen key to seal the sensor to the measurement chamber.

Note: Remove the inflow tube before disassembling the 2 or 5 mm sensors.

Figure 2 Installation of flow-through unit for NITRATAX clear sc

1

Sample drain

4

Label (pay close attention to clearance requirements)

2

Ø 8 mm thru, clearance hole

5

Sample inlet

3

Spacer (2x) (wrap sensor cable)