Table 2 calibration data of lck 325 formaldehyde, 4 check entered data – Hach-Lange POCKET Colorimeter II Application Instruction User Manual

Page 23

Advertising
background image

System start up

23

10. Press GREEN key. Enter "10.00" by using BLUE key to change value and

GREEN key to move to the next space. Press GREEN key until "A1" is

displayed.

11. Press GREEN key. Enter "0.993" by using BLUE key to change value and

GREEN key to move to the next space. Press GREEN key until "Add" is

displayed.

12. Press MENU until "Edit" is displayed. Press GREEN key until „S0“ is displayed.
13. Press GREEN key. Enter "00.50" by using BLUE key to change value and

GREEN key to move to the next space. Press GREEN key until "A0" is

displayed.

14. Press GREEN key. Enter "0.046" by using BLUE key to change value and

GREEN key to move to the next space. Press GREEN key until „S1“ is

displayed.

15. Press MENU until "Edit" is displayed to check the entered data. Press GREEN

key. „RES“ is displayed.

16. Press BLUE key to scroll through the entries. Compare the entries with the data

shown in the table below.

17. After comparing the entries press MENU two times to enter the measuring

mode. See

4.3 Evaluate LANGE Formaldehyde Cuvette Test on page 25

.

Note: In case of any mistake retrieve the factory calibration first. See

3.5.1 Retrieve the factory

calibration on page 24

. After retrieving the factory calibration, start the programming procedure

again. If the instrument is shut-off or loses power during data entry, all edits will be lost.
Automatic shut-off is 60 minutes.

3.4 Check entered data

To avoid erroneous measurements it is recommended to check the data after
completing the programme procedure.

1. Press MENU for about five seconds until "USER" followed by "CAL" is

displayed.

2. Press BLUE key to scroll to "Edit". Press GREEN key.
3. "RES" is displayed. Press BLUE key to scroll through the entries. Compare the

entries with the figures in the

3.2.2.1 Evaluation as Formaldehyde on page 22

.

In case a value was not stored correctly, retrieve the factory calibration first. See

3.5.1

Retrieve the factory calibration on page 24

. After retrieving the factory calibration,

start the programming procedure again.

Table 2 Calibration data of LCK 325 Formaldehyde

RES

--.--

S0

0.50

A0

0.046

S1

10.00

A1

0.993