Hach-Lange DR1900 Basic User Manual User Manual

Page 3

Advertising
background image

Table of contents

Specifications

on page 3

Startup

on page 10

General information

on page 4

Standard operation

on page 11

Installation

on page 6

Maintenance

on page 14

User interface and navigation

on page 8

Troubleshooting

on page 15

Additional information

Additional information is available on the manufacturer's website.

Specifications

Specifications are subject to change without notice.

Specification

Details

Measurement method

VIS spectrophotometer for HACH and HACH LANGE water analysis
methods

Measurement mode

Transmittance (%), Absorbance (Abs) and Concentration (Conc)

Dimensions (W x D x H)

178 x 261 x 98 mm (7.0 x 10.3 x 3.8 in.)

Enclosure rating

IP67 (closed sample cell cover)

Weight

1.5 kg (3.3 Ib)

Power requirements (internal)

AA size Alkaline (4x) or
rechargeable Nickel Metal Hydride (NiMH) batteries (4x) (optional module
necessary

1

)

Power requirements (external)

Power supply: 110–240 VAC; 50/60 Hz (optional module necessary

1

)

Interface

USB mini (optional module necessary

1

)

Operating temperature

10 to 40 °C (50 to 104 °F); maximum 80% relative humidity (non-
condensing)

Storage temperature

–30 to 60 °C (–30 to140 °F); maximum 80% relative humidity (non-
condensing)

Lamp source

Xenon flash

Wavelength range

340–800 nm

Photometric measuring range

±3.0 Abs (wavelength range 340–800 nm)

Wavelength accuracy

±2 nm (wavelength range 340–800 nm)

Spectral bandwidth

5 nm

Photometric accuracy

3 mAbs at 0.0 to 0.5 Abs, 1% at 0.50 to 2.0 Abs

Photometric linearity

< 0.5% to 2 Abs
≤ 1% at > 2 Abs with neutral glass at 546 nm

Wavelength selection

Automatic, based on the method selection

Stray light

< 0.5% T at 340 nm with NaNO

2

Repeatability

± 0.1 nm

Wavelength resolution

1 nm

English 3