3 lck 305 ammonium-nitrogen nh4-n, 3 lck 305 ammonium-nitrogen nh – Hach-Lange PCII-Ammonium LCK 303_304_305 User Manual

Page 13

Advertising
background image

System start up

13

14. Press GREEN key. Enter "0.044" by using BLUE key to change value and

GREEN key to move to the next space. Press GREEN key until „S1“ is
displayed.

15. Press MENU until "Edit" is displayed to check the entered data.

Press GREEN key. „RES“ is displayed.
Press BLUE key to scroll through the entries. Compare the entries with the data
shown in the table below.

16. After comparing the entries press MENU two times to enter the measuring

mode. See

4.3 Evaluate HACH-LANGE Ammonium-Nitrogen/ Ammonium

Cuvette Tests on page 17

.

Note: In case of any mistake retrieve the factory calibration first. See

3.5.1 Retrieve the factory

calibration on page 15

. After retrieving the factory calibration, start the programming procedure

again. If the instrument is shut-off or loses power during data entry, all edits will be lost.
Automatic shut-off is 60 minutes.

3.3.3

LCK 305 Ammonium-Nitrogen NH

4

-N

1.

Press POWER to switch on the instrument.

2.

Press MENU. "SEL" is displayed. A flashing arrow indicates the current
measuring range (1 or 2).

3.

Select measuring range 1 or 2 for LCK 305 (1-12 mg/l NH

4

-N).

Note: Press

GREEN

key to toggle between measuring ranges.

4.

Press MENU for about five seconds until "USER" followed by "CAL" is
displayed.

5.

Press BLUE key until "Edit" is displayed. Press GREEN key. " RES" is
displayed.

6.

Press GREEN key. "-.---" is displayed.

7.

Press BLUE key until "--.--" is displayed. Press GREEN key until "S0" is
displayed.

8.

Press GREEN key until "A0" is displayed.

9.

Press GREEN key until "S1" is displayed.

10. Press GREEN key. Enter "12.00" by using BLUE key to change value and

GREEN key to move to the next space. Press GREEN key until "A1" is
displayed.

11. Press GREEN key. Enter "1.871" by using BLUE key to change value and

GREEN key to move to the next space. Press GREEN key until "Add" is
displayed.

Table 4 Calibration data of LCK 304

RES

-.---

S0

0.015

A0

0.044

S1

2.000

A1

1.832