Specifications

Specifications are subject to change without notice.

Specification

Details

Range

0.000–5000 milli-Nephelometric Turbidity Units
(mNTU)
(0–5.0 NTU with +10% over-range limit)

Measurement units

mNTU, mFTU, NTU, FTU, mg/L, Degree

Over-range limit

*

+10%

Method compliance

USEPA Approved for drinking water applications by
Method 10133

Lowest expected reading

7 mNTU

Accuracy

±3% of reading or ±5 mNTU (whichever is greater)

Repeatability

±0.42% at 50 mNTU
±0.3% at 800 mNTU

Resolution

0.001 mNTU on lowest range
0.1 mNTU on highest range

Calibration

Single point at 800 mNTU ±50 mNTU

Calibration frequency

Every 3 months (or as per regulations)

Calibration (wet)
verification method

Pass/fail criteria or tolerance (standards less than
1000 mNTU): 25–50 mNTU
Pass/fail criteria or tolerance (standards greater than
1000 mNTU): 5%

Calibration (dry)
verification method

Verification level: 20–50 mNTU
Performance default: ±5 mNTU of the identified
baseline value
Method precision: less than or equal to 5% relative
standard deviation (RSD)

Signal averaging

No averaging, 6, 30, 60 or 90 seconds

Specification

Details

Sample light source

Class 1 Laser Product with embedded 7.5 mW,
670 nm, Class 3B Laser Source
Complies with 21 CFR 1040.10 except for deviations
pursuant to Laser Notice No. 50, dated June 24,
2007. FDA Laser Accession No. 9911570

Power requirement

10.5 to 13.2 VDC at 1.5 VA

Operating conditions

0 to 40 °C (32 to 100 °F), 5–95% non-condensing

Storage conditions

–20 to 60 °C (–4 to 140 °F)

Sample requirements

Temperature: 0 to 50 °C (32 to 121 °F)
Flow: 100–750 mL/minute (1.6–11.9 gal/hour)

Recorder output

0–20 mA or 4–20 mA
The output span is programmable over any portion of
the 0.000–1000 mNTU range.

Sample inlet fitting

¼-in. NPT female, ¼-in. compression fitting

Drain fitting

½-in. NPT female, ½-in. hose barb

Tubing requirements

Sample inlet: ¼-inch OD rigid or semi-rigid tubing
Drain: ½-inch ID flexible plastic tubing

Dimensions
(W x D x H)

25.4 x 30.5 x 40.6 cm (10 x 12 x 16 in.)

Mounting options

Wall mounted or floor stand

Shipping weight

Turbidimeter and controller—16.31 kg (13.5 lb)
Turbidimeter only—4.71 kg (10 lb)

ASTM conformance

D6698 Standard test method for online measurement
of turbidity below 5 NTU

Limit of detection (LOD)

0.296 mNTU (ISO 15839)

Protection class

III

Pollution degree

2/II