Hach-Lange 2100P iso - Portable Turbidimeter Instrument User Manual
Page 23

Section 2, continued
Page 23
4650098 operation.fm
Calibration
7. The display will show “S1” (with the 1 flashing) and 20 FNU or the value of the
S1 standard for the previous calibration. If the value is incorrect, edit using the
arrow keys.
8. Fill a clean sample cell to the line with well mixed 20 FNU StablCal Standard or
20 FNU formazin standard. Insert the sample cell into the cell compartment by
aligning the orientation mark on the cell with the mark on the front of the cell
compartment. Close the lid.
9. Press
READ
. The instrument will count from 60 to 0 (67 to 0 if signal average is
on). The display will automatically increment to the next standard.
10. Remove the sample cell from the cell compartment.
11. The display will show “S2” (with the 2 flashing) and “100 FNU” or the value of
the S2 standard for the previous calibration. If the value is incorrect, edit using
the arrow keys.
12. Fill a clean sample cell to the line with well mixed 100 FNU StablCal Standard
or 100 FNU formazin standard. Insert the sample cell into the cell compartment
by aligning the orientation mark on the cell with the mark on the front of the cell
compartment. Close the lid.
13. Press
READ
. The instrument will count from 60 to 0 (67 to 0 if signal average is
on). The display will automatically increment to the next standard.
14. Remove the sample cell from the cell compartment.
15. The display will show “S3” (with the 3 flashing) and “800 FNU” or the value of
the S3 standard for the previous calibration. If the value is incorrect, edit using
the arrow keys.
16. Fill a clean sample cell to the line with well mixed 800 FNU StablCal Standard
or 800 FNU formazin standard. Insert the sample cell into the cell compartment
by aligning the orientation mark on the cell with the mark on the front of the cell
compartment. Close the lid.