Diagnostic codes, Table 3 – Hach-Lange 2100 N IS Laboratory Turbidimeter Basic User Manual User Manual

Page 15

Advertising
background image

Note: Any calibration being calculated when an error occurs, is discarded. The old
calibration is kept.

Table 3 Error codes

Error

Description

Solution

ERR01 The turbidity of the

dilution water is greater
than 0.5 NTU.

Start the calibration again with lower turbidity
dilution water.

Note: Ignore ERR01 when the sample cell diameter
is less than 25 mm. Push UNITS/Exit to go back to
measurement mode.

ERR02

• Two calibration

standards have the
same value.

• The difference

between two
calibration standards
is less than
60.0 NTU.

• The turbidity of

Standard 1 is too
low (less than
10 NTU).

1. Inspect the preparation of standards.
2. Do the calibration again.

Note: Ignore ERR02 when the sample cell diameter
is less than 25 mm. Push UNITS/Exit to go back to
measurement mode.

ERR03 Low light error

1. Put the sample in the instrument again.
2. Make sure that the lamp light is on.
3. Make sure that an object is not in the light

path.

4. Do sample dilution if necessary.

ERR04 Memory malfunction

1. Turn the instrument off and then back on.
2. Contact Technical Support if the error

occurs again.

ERR05 A/D is over the range

1. Make sure that the light shield is closed.
2. Contact Customer Service if necessary.

Table 3 Error codes (continued)

Error

Description

Solution

ERR06 A/D is under the range

1. Make sure that no object is in the light

path.

2. Contact Customer Service if necessary.

ERR07 Light leak

1. Make sure that the cover for the sample

cell compartment is closed.

2. Turn the instrument off and then back on.

ERR09 Printer time out error

1. Make sure that the external printer is

connected correctly.

2. Make sure that the external printer is

selected (online).

ERR10 System voltage out of

range

1. Turn the instrument off and then back on.
2. Contact Customer Service if the error

occurs again.

ERR11 System loop test error

1. Turn the instrument off and then back on.
2. Contact Customer Service if the error

occurs again.

Diagnostic codes

Table 4

lists the diagnostic codes that are used to get information about

instrument operation when instrument operation is in doubt.
To do a diagnostic test:

1. Push SETUP.
2. Use the arrow keys to enter a diagnostic code.
3. Push ENTER to show the diagnostic value.
4. Push UNITS/Exit to go back to Measurement mode.