Ufield correction – Liquid Controls T350N-P User Manual
Page 7
Sponsler, Inc.
T350N-P Truck Totalizer/Printer
pg. 7
Doc# MN-T350N-P
For higher resolution (reading in gallons and tenths, gallons and hundredths, 
etc.) multiply by 10 for tenths; by one for hundredths. The number dialed into 
dipswitch S5 corresponds to the number of digits to the right of the decimal point. 
 
Example: 
 
 
Flowmeter “K” Factor = 230
CF= .004348
Desired display = gallons and tenths
Multiply by = .04348
S4 = 1
Factor Switches = 4348; S3 = 4, S2 = 3, S1 = 4, S0 = 8
Decimal point = S5 = 1
To adjust the calibration factor to reflect the turbines actual response to the 
operating conditions apply the following formula: 
Field Correction
 
 
New Calibration Factor = Actual Total
Meter Total
X Calibration Factor
 
Example #2: 
 
Actual Total = 50
Meter Total = 52
C.F. = .004348
50/52 X .004348 = New Calibration Factor
.9615 X .004348 = .0041806
Insert 4181 into S3 – S0 respectively
 
In the above example .9615 denotes that the meter is operating .385% fast and 
multiplying the present calibration factor (.004348) by the ratio of Actual Total: 
Meter Total (.9615) reduces the calibration factor 3.85%. 
 
Assume that rather than gallons, liters are the desired engineering unit.
Calibration Factor – Change of Calibration Engineering Units
 
Example #3: 
 
K-Factor = 230 pulses per gallon
Liters = 3.785 per gallon
 
 
3.785
230
= Calibration Factor
 
 
.01646 = Calibration Factor for display of liters