FUNCTION SELECTION
 
 
Reset Function 
Reset is valid only to load a new calibration factor. 
Reset is initiated externally by placing a magnet in the proximity of reset S6 located on the Display 
P.C.B. and indicated on the enclosure label as ‘RESET’. 
 
 
FIGURE 4
 
 
 
 
 
 
 
 
 
 
MISCELLANEOUS
 
 
Factored Digital Pulse Output 
 
This signal output is an open drain configuration that can sink 1A continuously. The maximum drain 
voltage should not exceed 150VDC (V+). A pull-up resistor must be in series with V+ and TB1-3. The 
output is a 2ms negative pulse for each engineering unit i.e. 1 pulse for each gallon, lb., etc.. 
 
 
Typical Configuration-
 
 
 
 
 
IT200M
SIGNAL OUT +
SIGNAL OUT –
(V-)
V+
TB1-3
TB1-4
RL