Luminex MasterPlex CT Tutorial Manual User Manual

Page 6

Advertising
background image

http://www.miraibio.com

Plex

CT

6

Thumbwheel

Bolt

The initial Adjust Probe window gives detailed instructions on adjusting the probe

height. To adjust the vertical height of the sample probe:

1. Remove the clear plastic shield that covers the sample probe area.

2. Press the Eject button and place a plate in the plate holder.

3. Place the appropriate alignment tool in the plate:

ƒ For a standard plate with flat-bottom wells – stack two of the larger

(5.08 mm diameter) alignment discs together and place them into a

selected well

ƒ For a filter bottom plate – stack three of the larger (5.08 mm diameter)

alignment discs together and place them into a selected well

ƒ For a half-volume plate with flat-bottom wells – stack two of the

smaller (3.35 mm diameter) alignment discs together and place them into

the selected well

ƒ For a round-bottom (U-bottom) plate – stack two of the smaller (3.35

mm diameter) alignment discs and place them into a selected well

ƒ For a plate with conical wells – place one alignment sphere into a

selected well