Wiring diagrams – Meriam M1500 Digital Pressure Transmitter User Manual

Page 17

Advertising
background image

13

Wiring Diagrams

Wiring Diagrams

Wiring Diagrams

Wiring Diagrams

Analog Wiring Diagrams

Analog Wiring Diagrams

Analog Wiring Diagrams

Analog Wiring Diagrams


2-wire 4-20 mA, Loop Powered

24 VDC nominal – see “Power Requirements” in the Specifications section of this manual for more detail
Note: Switch outputs are not available in 2-wire 4-20 mA service.

4-wire 4-20mA with Switches

24 VDC nominal – lsolated loop power supply

12 VDC nominal – External power supply

see “Power Requirements” in the Specifications section of this manual for more details

Can be placed
on either side of
the loop.

Not Available
for 2-wire 4-20
mA service

.

4-wire 4-20mA with Switches, con’t

1. For 4-wire commissioning always connect external supply to PWR and GND terminals before
connecting isolated loop supply to LP+ and LP–.
2. Isolated power supply is required for LP+ / LP– loop.
3. Do not short LP– and GND terminals.
4. Switches are solid state, opto-couplers with SPST operation (Clare model CPC1008N). Switch
circuits are isolated from each other and from the M1500 mA output.

Advertising
This manual is related to the following products: