Meriam M130 T_C Thermocouple Calibrator User Manual

Page 15

Advertising
background image

M130 – T/C Calibrator

Instruction Manual Z1017-IR

E.O. 6566 Sept. 2008

www.meriam.com

Page 15 of 21

session will stop and the RECORD Data mode will automatically exit to
Measure Mode.

VIEW Recorded Data:
To view the results of a REC session from Measure Mode, press SETUP key
and then W or X key until VIEW is displayed. Press ENTER. The REC icon
is shown in upper right. Display will show recorded value 1 of X (1/X).
Pressing ENTER will toggle to show the T/C type and engineering unit. The
W or X key moves to previous or next recorded value, respectively.


OFF Timer

To help conserve battery life, the M130 can be programmed to shut off
automatically after 30 minutes of keypad inactivity. From Measure Mode,
press SETUP and then W or X key until OFF TMR is displayed. Press the
ENTER key and use + or – key to display OFF or 30 MIN from menu. Press
ENTER key to save selection. Press SETUP again to return to Measure Mode.


Zeroing the M130

Preparation: Allow the M130 and supplied Shorting Plug (p/n Z9A83) to soak
at room temperature for 30 minutes in a controlled temperature environment.
Record the room temperature immediately prior to zeroing.


Zeroing: From Measure Mode, press SETUP once and then the W or X key
until F RECAL is displayed. Press the ENTER key to display ZERO. Press
ENTER to start ZERO process. Display will show SHORT. Install the
Shorting Plug (white, mini-T/C connector with internal copper short) on the
M130’s connection and press ENTER key. Display will show BUSY.

When ZERO is successfully completed, the message “ZERO PASSED” is
briefly displayed and the M130 is automatically returned to the Measure
Mode. Remove the Shorting Plug. The new temperature reference for the
M130 will be the ambient temperature recorded prior to zeroing. (Note:
factory reference temperature is 23

°C.)


If the ZERO is unsuccessful, the message “ZERO FAILED” will be displayed
briefly. In this case, check to be sure the Shorting Plug is properly installed
and repeat the ZERO process. If “ZERO FAILED” appears again contact
Meriam for technical support.