Meriam MANOMETERS User Manual

Page 2

Advertising
background image

Density is a function of temperature and gravity is

a function of latitude and elevation. Because of this

relationship some ambient conditions must be selected as

standard so that pressure bears a fixed definition.

Standard conditions for mercury used as a unit of pressure:

Gravity: 980.665 cm/sec

2

(32.174 ft/sec

2

)

at sea level and 45.544 degrees latitude

Temperature: 0°C (32°F) density = 13.5951 g/cm

3

Standard conditions for water used as a unit of pressure.

Gravity: 980.665 cm/sec

2

(32.174 ft/sec

2

)

at sea level and 45.544 degrees latitude

Temperature: 4°C (39.2°F) density = 1 g/cm

3

Though it is recommended that the value of a water

column as a unit of pressure be at 4°C, its universal

acceptance has been slow. For instance in aeronautics

15°C (59°F) is used. The American Gas Association uses

15.56°C (60°F), and in orifice flowmeter work 20°C

(68°F) is commonly used.

Recognizing a manometer may be read outside, standard

temperature and gravity corrections can be applied to

improve the accuracy of a manometer reading at any given

conditions.

Fluid Density Corrections

Manometers indicate the correct pressure at only one

temperature. This is due to the fact that the indicating

fluid density changes with temperature. If water is the

indicating fluid, an inch scale indicates one inch of water

at 4°C only. On the same scale mercury indicates one

inch of mercury at 0°C only. If a reading using water or

mercury is taken at 20°C (68°F) then the reading is not an

accurate reading. The error introduced is about 0.4% of

reading for mercury and about 0.2% of reading for water.

Since most manometers are read at temperatures well

above the standard temperature, corrections are needed.

A simple way of correcting for the temperature error is to

ratio the densities.

(Standard) ρ

o

gh

o

= (Ambient) ρ

t

gh

t

h

o

= the corrected height of the indicating fluid to standard

temperature

h

t

= height of the indicating fluid at the temperature

when read

ρ

o

= density of the indicating fluid at standard temperature

ρ

t

= density of the indicating fluid at the temperature

when read

6. Reading

For consistent results, it is necessary that the fluid

meniscus always be observed in the same way. A

convex meniscus forms when mercury is used. In

this instance the fluid level should be observed from

the upper most point. For all other indicating fluids

a concave meniscus forms. The reading in this case

should be observed from the lowest point of the

meniscus. To duplicate factory calibration of incline

manometers, this technique needs to be followed.

It is important to remember the levels in both legs of

U-tube manometers must be read and these readings

added together to obtain an actual indication.

A plane tangent to the fluid meniscus and at a right

angle to the tube bore intercepts the scale where it

should be read.

7. Theory

The fundamental relationship for pressure expressed

by a liquid column is:

p = P

2

- P

1

= ρgh

p = differential pressure

P

1

= pressure applied to one liquid surface

P

2

= pressure applied to the other liquid surface

ρ

= mass density of the liquid (specific gravity)

g = acceleration of gravity

h = height of the liquid column

In the case of absolute manometers (barometers), P

1

is equal to zero absolute pressure, simplifying the

equation to:

P = ρgh

As simple as manometry is, certain aspects are often

overlooked. Manometry incorporates both a value for

density and gravity. These two values are not constant.

Advertising
This manual is related to the following products: