6 appendix, 1 technical specifications – Multichannel Systems MEA-IT Manual User Manual
Page 29

25
6 Appendix
6.1 Technical
Specifications
The MEA-IT-System is a device for impedance testing of microelectrode arrays.
Warning: The device may only be used together with microelectrode arrays from
Multi Channel Systems MCS GmbH, and only for the specified purpose. Damage of
the device and even fatal injuries can result from improper use. Do not open the data
acquisition box and do not change hardware configuration as it could lead to improper
behavior of the system.
General characteristics
Operating temperature
10 °C to 50 °C
Storage temperature
0 °C to 50 °C
Relative humidity
10 % to 85 %, non-condensing
Dimensions (W x D x H)
250 mm x 140 mm x 25 mm
Weight
1100 g
Number of channels
60
Measurement range
5 k
to 2 M
Test signal
100 mV, 1 kHz Sinus
Accuracy 5
%
Power supply
USB powered
PC interface
USB 2.0
Software
Operating system
Microsoft Windows ® 7 (32 or 64 bit), XP or Vista
with NTFS, English and German version supported
MEA-IT program
Version 1.0.0 and higher