Connection of the perfusion cannula ph01 – Multichannel Systems MEA2100-System User Manual
Page 4

6
1. Connect the PPS2 with the power supply unit to the power outlet.
2. Connect the PPS2 with the USB cable A-B to the computer for PPS2 software control (possible
via an USB hub).
3. Connect the tubes of the PPS2 to the perfusion solution, the PH01, the perfusion out and the waste
bottle. Please follow the instructions on the leaflet in the tube bag. Adjust the flow rate.
Power Supply
Unit
USB Cable A-B
PPS2 Rear Panel
Perfusion In
(via PH01)
Perfusion Out (from MEA)
Waste Bottle
Perfusion Solution
Connection of the Peristaltic Perfusion System PPS2
Power Outlet
5
TC02 Rear Panel
Connection of the Perfusion Cannula PH01
Perfusion Cannula PH01
CA Cable: D-Sub9 to Pt100
1. Connect the perfusion cannula PH01 with the CA cable (D-Sub9 to Pt100) to the temperature
controller TC02. The connector is symetrical, so the orientation of the connector is irrevlevant.
2. Set the respecitve channel of the temperature controller via control software or via touch screen to the
device PH01. Please see the TCX manual (CD: Manuals/MEA-System/TC01-02_Manual_RevG.pdf) for
detailed information.