4.3 MCS Default PI Coefficients

Note: The following PI parameters have been optimized at an ambient temperature of 25 °C,
the PI coefficients for use with the PH01 at a flow rate of 3 ml/min. You may have to adjust these
PI coefficients for your experimental setup, especially if the ambient temperature or the flow rate
differ at large from those used by MCS. Using suboptimal PI coefficients can lead to an oscillation
of the actual temperature, which is harmless, but may result in an unwanted behavior of the
temperature controller.

TCW1

PH01

MEA 1060-UP

MEA 1060-INV

P 60.00 2.00 4.00

6.00

I 75.00 2.00 1.50

0.90

4.4 Technical

Specifications

Operating temperature

10 °C to 40 °C

Storage temperature

0 °C to 50 °C

Dimensions (W x D x H)

170 mm x 224 mm x 66 mm

Weight

1.5 kg

Supply voltage and current

24 V and 4 A

Desktop AC Power Adapter

85 VAC to 264 VAC @ 47 HZ to 63 Hz

Sensor type

PT 100

Measuring method

four wire measuring bridge

Measuring temperature range

0 °C to 105 °C

Number of output channels

1 (TC01), 2 (TC02)

Output voltage

max. 24 V

Output current

max. 2.5 A per channel

Output power

max. 30 W per channel

Resistance of heating element

5 - 100

Control range

Ambient temperature (min. 5 °C) to 105 °C

Control interface

USB 2.0

Note: Outputs are isolated against ground.