Multichannel Systems 256MEA-9well-STIM-ADPT User Manual

Multi channel systems, 256mea-9well-stim-adpt

Advertising
background image

Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany

Fon +49-7121-9 09 25- 0
Fax +49-7121-9 09 25-11

[email protected]
www.multichannelsystems.com

© 201

2 Multi Channel Systems MCS GmbH

Product information is subject to change
without notice.

256MEA-9well-STIM-ADPT

Adapter for Electrode Stimulation of 9-Well MEAs in USB-MEA256-Systems

multi

channel

systems

*

Please fix the adapter 256MEA-9well-STIM-ADPT on the USB-MEA256 amplifier as shown on the foto.
Connect the red connectors (2 mm) with the provided cables to the STG for stimulation of S1 and S2.
The black connectors (2 mm) are for grounding the system.

Stimulation of S1 means, that the S1 electrodes in each of the nine wells are stimulated altogether.

Stimulation of S2 means, that the S2 electrodes in each of the nine wells are stimulated altogether.

It is not possible to stimulate only one of the nine stimulation electrodes S1 or S2, for example.

Rear panel of the adapter

Front panel of the adapter
256MEA-9well-STIM-ADPT.
The red connectors are for
stimulation of the electrodes
S1 and / or S2 in each well.
The black connectors are
for grounding the system.

Example Well A : The numbering of MEA electrodes in the
6 x 5 grid per each well follows the standard numbering scheme
for square grids: The first digit is the column number and the
second digit is the row number. For example, electrode 23 is
positioned in the second column of the third row. Two square
electrodes (S1 and S2) per well are available for stimulation or
recording and additionally one big internal reference electrode.

A12 A22 A32 A42 A52 A62

A

A21 A31 A41 A51

A13 A23 A33 A43 A53 A63

A14 A24 A34 A44 A54 A64

A25 A35 A45 A55

A
S
1

A
S
2

Pin layout
9-Well MEAs

Warning: The 256MEA-9well-STIM-ADPT may only be used together with the
USB-MEA256-System from Multi Channel Systems MCS GmbH, and only for the
specified purpose. Damage of the device and even injuries can result from improper use.