4.9.5 Triggering on Extracted Parameters
Similar to the triggering on raw data streams, you can also define a threshold for an extracted
parameter stream to generate trigger events. For example, you can trigger the recording only
if the spike rate reaches a threshold level.
Select the appropriate parameter data stream and channel from the Channels list.
If you select the option Negative Slope, the fall of the parameter curve is regarded.
If you select Positive Slope, the rise is considered.
If you use a Parameter stream with multiple extracted parameters you have to specify
the parameter type as well.
4.9.6 Time Based Trigger
You also can define a time-based trigger. The trigger event can either occur periodically
or at specified points of time. For example, you can use this feature if you want to apply
a drug at specific time intervals and want to record data only at the drug application times.