0 installation – Nova-Tech Density Kit for Adam PW and PGW Series User Manual

Page 5

Advertising
background image

4.0 INSTALLATION

• Replace the original pan with the pan and hanger (7) supplied

with the kit.

NOTE:

For the kit supplied for PGW, the pan with hanger (7)

needs to be assembled. Remove the two nuts from the hanger
ends, insert the two ends through the holes of the substitute pan
from top and fasten the nuts at the bottom of the pan to keep the
hanger in the upright position.

• Place the beaker support (5) over the hanger so that they do not

touch each other.

• Fill the beaker (6) with either distilled water at room temperature,

alcohol or another liquid of known density. For determination of a
liquid’s density add the liquid to the beaker. NOTE: The liquid
used to fill the beaker must cover the bottom pan of the sample
pan (8) as well as the solid item being put in this pan.

• Fix the thermometer clip (9) on the beaker wall and insert the

thermometer (4) into the liquid.

• Hang the sample pan (8) on the hanger (7) if the kit is to be used

for density determination of solid.

• Hang the glass float (1) on the hanger (7) when the kit is being

used for the density determination of liquid.

PGW with the Density Kit installed

© Adam Equipment Company 2007

3