Calibration

112

Series N5700 User’s Guide

*OPC?

Step 4. Measure the output voltage with the Agilent 3458A and enter the

data.

CAL:DATA <data>

Step 5. Select the second voltage calibration point.

CAL:LEV P2

*OPC?

Step 6. Measure the output voltage with the Agilent 3458A and enter the

data.

CAL:DATA <data>

Current Programming and Measurement Calibration

Step 1. Connect a precision shunt resistor to an output. Connect the Agilent

3458A to measure the voltage across the shunt. The shunt resistor
should be suitable for measuring at least 120% of the power supply’s
rated full-scale current.

Step 2. Select the full-scale current programming range. The value to

program a range must be the maximum current of the range.

CAL:CURR <current>

Step 3. Select the first current calibration point.

CAL:LEV P1

*OPC?

Step 4. Calculate the shunt current (I=V/R) and enter the data.

CAL:DATA <data>

Step 5. Select the second current calibration point.

CAL:LEV P2

*OPC?

Step 6. Calculate the shunt current (I=V/R) and enter the data.

CAL:DATA <data>

Exit Calibration mode

CAL:STAT OFF