GASS Module User Manual

4

Bulletin #217

1. Introduction

Perma Pure GASS Module™ sample conditioning systems are designed to prepare gas samples for
analysis. The GASS Module system will remove, particulates, mists and water vapor without removing
compounds being monitored.

A heated sample line is connected to the top fitting on the GASS Module (see Figure 1). Sample gas flows
to the filter where particulates are removed and/or mists are coalesced. After filtration, sample is passed
through the Nafion® Dryer where water vapor is selectively removed. Sample is now ready for analysis.

Standard Features Include:

•

Filter - Particulate and/or Coalescing

•

Perma Pure Nafion

®

Gas Dryer

•

Purge Eductor

•

Filter Thermostat

•

Dryer Heater Thermocouple

•

Module Housing

•

Mounting Clamps

Figure 1

Options:

•

Automatic Filter Drain

•

Filter Heater Thermocouple and Extension Wire