Calibration Gas Usage Configuration


“Calibration Gas Usage Configuration” is a screen (Figure 20) to associate the analyzers
configured in the “Analog Input Configuration” screen and the gases configured in the
“Calibration Gas Configuration” screen. This screen will determine the gas used to calibrate each
analyzer.


Open this screen from the Main Menu by selecting the ‘Configuration’ menu item and then
selecting ‘Cal Gas Usage’ from the submenu.

Figure 20

This screen will display all configured analyzers versus all configured cal gas to allow the
selection of specific gases for the different calibrations of the analyzers, Zero, Low, Mid, High
and Span.